Reporting guide of unique device identification database
1 Scope
This standard specifies the basic requirements for reporting device identifier and related information to the unique device identification database (hereinafter referred to as the database).
This standard is applicable to reporting of unique device identification database.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
YY/T 1752-2020 Basic data set of unique device identification database
3 Terms and definitions
For the purposes of this standard, the terms and definitions established in YY/T 1752-2020 as well as the followings apply.
3.1
data interface
interface between computer software systems to transfer data and exchange information, which is realized in the form of electronic files.
4 Abbreviations
For the purposes of this document, the following abbreviations apply.
XML: Extensible Markup Language
JSON: JavaScript Object Notation
5 Data reporting mode
For the reporting, webpage, data template import and data interface exchange may be used.
When the webpage mode is adopted, the relevant data shall be reported according to the webpage reporting tips. See Annex A for the contents of data reporting.
When the data template import mode is adopted, the data shall be reported according to the data template provided by the unique device identification database, and the data format and content shall meet the relevant requirements of the data template before importing into the database. See Annex A for the contents of data reporting.
When the data interface exchange mode is adopted, the data interaction interface shall be developed, and the data exchange relationship may be established with the database after registration of docking services through the database. The exchange format of data interface may be XML and JSON. See Annex B for the examples of formats.
Foreword i
1 Scope
2 Normative references
3 Terms and definitions
4 Abbreviations
5 Data reporting mode
6 Data interface exchange format requirements
7 Data reporting content
8 Security requirements for data reporting
Annex A (Informative) Examples for reporting of basic information of the unique device identification database
Annex B (Informative) Examples of interfaces for reporting of the unique device identification database